Verification and Comparison of Sysmex KX-21N, Cell-Dyn 1800 and Horiba ABX Micros ES 60 Automated Hematology Analyzers at Tikur Anbessa Specialized Hospital Addis Ababa, Ethiopia
No Thumbnail Available
Date
2017-06
Authors
Journal Title
Journal ISSN
Volume Title
Publisher
Addis Ababa University
Abstract
Background: Laboratory testing plays a crucial role in the detection, diagnosis and treatment ofdisease in patients. Analytical evaluation of a new analyzer or method should be done in order toconfirm the declared specifications before its introduction into the routine use. Agreement of allautomated hematology analyzers must be carried out in all laboratories. But, the practice of usingautomated analyzers interchangeably without evaluation is common in our country includingTikur Anbessa specialized hospital (TASH).
Objective: To verify performance and compare agreement of Cell-Dyn 1800, Sysmex KX-21Nand Horiba ABX micros ES 60 hematology analyzers found in TASH.
Methods: A cross sectional study was conducted to verify performance and assess agreementbetween Cell-Dyn 1800, Sysmex KX-21N and Horiba ABX micros ES60hematology analyzersat Tikur Anbessa specialized hospital, Addis Ababa, Ethiopia. The study period was fromFebruary to May 2017. According to the CLSI guideline a total of 267 samples were analyzed induplicates on the three analyzers to verify and compare them. Data cleaning and analysis wereperformed by Microsoft excel 2007 and SPSS version 20.
Result: We found that minimal carryover (< 0.5%) and excellent linearity for white blood cell,hemoglobin, red blood cell and platelet (r > 0.999). Precision was good for all major CBCparameters with % CV< 2.0. Agreement between most of the parameters of the analyzers lies inthe limits of agreement for more than 95% of cases according to the Bland-Altman plot analysis.Pearson correlation and linear regression graph showed that the three analyzers were wellcorrelated (r > 0.900) for major CBC parameters. The correlation of WBC differential-neutrophil and lymphocyte were excellent (r > 0.920). Poor correlation was observed in mixedcells (r=0.005) especially in the high levels between Horiba and Sysmex. There was alsomoderate correlation in mean platelet volume (MPV) between analyzers of Horiba and Sysmex(r= 0.593 and r=0.572) at normal and high levels respectively.
Conclusion: The overall performance of the three analyzers is excellent. They agree well witheach other and can be used interchangeably with care in mixed cells and MPV.
Key words: Method comparison, performance evaluation, verification of hematology analyzersSysmex KX-21, Horiba ABX micros ES 60, Cell-Dyn 1800, hematology