Comparision of Hemocue301+ With Sysmex-Kx21n and Celldyn1800 Automated Hematology Analyzers for Measurement of Hemoglobin Levels in Patients at Stt. Paul’s Hospital Millennium Medical College, Addis Ababa, Ethiopia, 2015
No Thumbnail Available
Date
2016-10
Authors
Journal Title
Journal ISSN
Volume Title
Publisher
Addis Ababa University
Abstract
Background:Hemoglobin concentration is measured routinely using automated hematology
analyzers. Even if these automated hematology analyzers are very accurate and reliable, they are
expensive and transport of the samples to the laboratory delays the process which may delay
treatment. Point of care hemoglobin analyzer like Hemocue301+ has potential to solve such
problems. Although method comparison experiment has to be conducted when new instrument
are installed, this practice is not common in our country.
Objective:To compare hemoglobin determined by Hemocue301+, with Sysmex-kx21N and
Cell-Dyn1800 methods using patients’ blood sample at St Paul’s Specialized Hospital
Millennium Medical College in Addis Ababa, Ethiopia.
Methods:A facility based cross sectional study with convenient sampling method was conducted
at St. Paul’s specialized teaching hospital starting from May 15 to June 2 2015. A total of
117specimens for agreement study and 3 specimens for precision study were analyzed on
Hemocue301+, Sysmex-Kx21N, and cell-Dyn1800 hematology analyzers to compare their
Hemoglobin results. This study includes patients of all age group. One sample t-test and bland
and Altman plot with limit of agreement were used to make inference. Data analyses were
performed using SPSS version 20 Software.
Result:The precision experiment showed within run normal range CV% for Hemocue301+,
SysmexKx21N and Cell-Dyn1800 (1.7%, 1.3%, 1.8%, respectively). The one sample t-test
showed a non-significant p-value of 0.10 for Hemocue301+ and Sysmex-Kx21N and the limit of
agreement from Bland and Altman plot for 95% was (-1.07, 1.25)g/dl with only 1.7% of the
cases were out of the limit of agreement. For Hemocue301+ and Cell-Dyn1800, a non-significant
p- value of 0.46 and the Bland and Altman plot limit of agreement for 95% was (-1.22, 1.31)g/dl
with only 1.7%of the case were out of the limit of agreement.Hemocue301+was agreed with
SysmexKx21N and Cell-Dyn1800 within 1 g/dl which was considered as clinically important
limit for Hemoglobin measurement.
Conclusion: The Hemocue301+ maybe used interchangeably with Sysmex-Kx21N and Cell-
Dyn1800 automated Hematology analyzers in St Paul’s Specialized Hospital Millennium
Medical College Hematology laboratory