Application of Graphite Based Solid State Iodide Selective Electrode for the Determination of Iodine in Iodized Salts and Ascorbic Acid

dc.contributor.advisorMoges, Ghirma (PhD)
dc.contributor.authorMequanint, Tiruwork
dc.date.accessioned2018-06-26T08:20:37Z
dc.date.accessioned2023-11-09T16:17:37Z
dc.date.available2018-06-26T08:20:37Z
dc.date.available2023-11-09T16:17:37Z
dc.date.issued2009-07
dc.description.abstractA laboratory-made iodide selective solid-state electrode (Ag2S-AgI) was prepared. The electrode response was linear in the iodide concentration range of 10-6 M to 10-1 M, with a slope of 56.85 mV/decade and detection limit of 6x10-7 M iodide. The application of the laboratory-made iodide electrode (vs. a saturated calomel reference electrode) was investigated for direct potentiometric titration of iodized salts and for potentiometric back titration of vitamin C in pharmaceutical products and orange juices. The electrode was used as an endpoint indicator system in place of the traditional starch-triiodide complex. Vitamin C with I3- was back titrated against Na2S2O3 in acidic media while acidified samples of table salt iodized with iodate (in presence of excess concentration of iodide) was directly titrated with Na2S2O3. The curves of both titrations exhibited distinct endpoints down to 10 ppm as potassium iodateen_US
dc.identifier.urihttp://etd.aau.edu.et/handle/12345678/3614
dc.language.isoenen_US
dc.publisherAddis Ababa Universtyen_US
dc.subjectSalts and Ascorbic Aciden_US
dc.titleApplication of Graphite Based Solid State Iodide Selective Electrode for the Determination of Iodine in Iodized Salts and Ascorbic Aciden_US
dc.typeThesisen_US

Files

Original bundle
Now showing 1 - 1 of 1
No Thumbnail Available
Name:
Tiruwork Mequanint.pdf
Size:
719.47 KB
Format:
Adobe Portable Document Format
License bundle
Now showing 1 - 1 of 1
No Thumbnail Available
Name:
license.txt
Size:
1.71 KB
Format:
Plain Text
Description:

Collections